Abstract
Classical pharmacokinetic models used in computer-controlled infusion pumps (CCIPs) assume instantaneous mixing of drug in blood; however, the average recirculation time of blood in man is approximately one minute. To investigate the effects of recirculation dynamics on the transient performance of CCIPs, we propose a hybrid physiologically-based pharmacokinetic model for the narcotic alfentanil. A three-compartment model was derived from the response of the hybrid model to a short infusion and used to compute a CCIP infusion targeting 450 micrograms/l. For this infusion, the hybrid model predicts that the arterial plasma concentration will overshoot the target concentration by 39 percent with an average prediction error of 3 percent. The overshoot and average prediction error increase to 100 and 25 percent respectively when using a three-compartment pharmacokinetic model derived from a bolus. The overshoot can be reduced by decreasing the maximum possible infusion rate, or by increasing the zero-order hold infusion interval.Keywords
This publication has 22 references indexed in Scilit:
- Noninvasive Cardiac Output MeasurementAnesthesia & Analgesia, 1992
- A simple analytical solution to the three-compartment pharmacokinetic model suitable for computer-controlled infusion pumpsIEEE Transactions on Biomedical Engineering, 1991
- Pharmacokinetic Model-driven Infusion of FentanylAnesthesiology, 1990
- Pharmacokinetics of Fentanyl Administered by Computer-controlled Infusion PumpAnesthesiology, 1990
- The Prospective Use of Population Pharmacokinetics in a Computer-Driven Infusion System for AlfentanilAnesthesiology, 1990
- Algorithm for optimal linear model-based control with application to pharmacokinetic model-driven drug deliveryIEEE Transactions on Biomedical Engineering, 1990
- Total intravenous anaesthesia with propofol and alfentanil by computer‐assisted infusionAnaesthesia, 1988
- Population Pharmacokinetics of AlfentanilAnesthesiology, 1987
- Plasma Concentrations of Alfentanil Required to Supplement Nitrous Oxide Anesthesia for General SurgeryAnesthesiology, 1986
- Potential pitfalls in the conventional pharmacokinetic studies: Effects of the initial mixing of drug in blood and the pulmonary first-pass eliminationJournal of Pharmacokinetics and Biopharmaceutics, 1979