Precise and Traceable 13C/12C Isotope Amount Ratios by Multicollector ICPMS
- 20 June 2008
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 80 (15) , 5963-5969
- https://doi.org/10.1021/ac800621u
Abstract
A new method for the measurement of SI traceable carbon isotope amount ratios using a multicollector inductively coupled mass spectrometer (MC-ICPMS) is reported for the first time. Carbon (13)C/(12)C isotope amount ratios have been measured for four reference materials with carbon isotope amount ratios ranging from 0.010659 (delta(13)C(VPDB) = -46.6 per thousand) to 0.011601 (delta(13)C(VPDB) = +37 per thousand). Internal normalization by measuring boron (11)B/(10)B isotope amount ratios has been used to correct for the effects of instrumental mass bias. Absolute (13)C/(12)C ratios have been measured and corrected for instrumental mass bias and full uncertainty budgets have been calculated using the Kragten approach. Corrected (13)C/(12)C ratios for NIST RM8545 (Lithium Carbonate LSVEC), NIST RM8573 (L-Glutamic Acid USGS40), NIST RM8542 (IAEA-CH6 Sucrose) and NIST RM8574 (L-Glutamic Acid USGS41) differed from reference values by 0.06-0.20%. Excellent linear correlation (R = 0.9997) was obtained between corrected carbon isotope amount ratios and expected carbon isotope amount ratios of the four chosen NIST RMs. The method has proved to be linear within this range (from (13)C/(12)C = 0.010659 to (13)C/(12)C =0.011601), and therefore, it is suitable for the measurement of carbon isotope amount ratios within the natural range of variation of organic carbon compounds, carbonates, elemental carbon, carbon monoxide, and carbon dioxide. In addition, a CO2 gas sample previously characterized in-house by conventional dual inlet isotope ratio mass spectrometry has been analyzed and excellent agreement has been found between the carbon isotope amount ratio value measured by MC-ICPMS and the IRMS measurements. Absolute values for carbon isotope amount ratios traceable to the SI are given for each NIST RM, and the combined uncertainty budget (including instrumental error and each parameter contributing to Russell expression for mass bias correction) has been found to be < 0.1% for the four materials. The advantage of the method versus conventional gas source isotope ratio mass spectrometry measurements is that carbon isotope amount ratios are measured as C(+) instead of CO2(+), and therefore, an oxygen (17)O correction due to the presence of (12)C(17)O(16)O(+) is not required. Organic compounds in solution can be measured without previous derivatization, combustion steps, or both, thus making the process simple. The novel methodology opens new avenues for the measurement of absolute carbon isotope amount ratios in a wide range of samples.Keywords
This publication has 32 references indexed in Scilit:
- Reformulated 17O correction of mass spectrometric stable isotope measurements in carbon dioxide and a critical appraisal of historic ‘absolute’ carbon and oxygen isotope ratiosGeochimica et Cosmochimica Acta, 2008
- Preparation of Synthetic Isotope Mixtures for the calibration of carbon and oxygen isotope ratio measurements (in carbon dioxide) to the SIInternational Journal of Mass Spectrometry, 2007
- Determination of lead isotope ratios in crude oils with Q-ICP/MSJournal of Analytical Atomic Spectrometry, 2007
- Determination of Pu isotope concentrations and isotope ratio by inductively coupled plasma mass spectrometry: a review of analytical methodologyJournal of Analytical Atomic Spectrometry, 2007
- Forensic applications of isotope ratio mass spectrometry—A reviewForensic Science International, 2006
- Referencing strategies and techniques in stable isotope ratio analysisRapid Communications in Mass Spectrometry, 2001
- A comparison between quadrupole, double focusing and multicollector ICP-MS instruments : Part I. Evaluation of total combined uncertainty for lead isotope ratio measurementsJournal of Analytical Atomic Spectrometry, 2001
- The chemical preparation and characterization of specimens for "absolute" measurements of the molar mass of an element, exemplified by silicon, for redeterminations of the Avogadro constantMetrologia, 1995
- Mass spectrometry of inductively coupled plasmasAnalytical Chemistry, 1986
- Isotopic standards for carbon and oxygen and correction factors for mass-spectrometric analysis of carbon dioxideGeochimica et Cosmochimica Acta, 1957