Comparison of Resting Hemodynamic Indices and Exercise Performance During Atrial Synchronized and Asynchronous Ventricular Pacing
- 1 March 1983
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 6 (2) , 202-209
- https://doi.org/10.1111/j.1540-8159.1983.tb04347.x
Abstract
Resting hemodynamic indices and exercise tolerance were measured during atrial synchronized (VAT) and asynchronous ventricular pacing (VOO) in 35 patients with implanted pacemakers which could be externally programmed to function in either pacing mode. Cardiac output and mean systemic arterial pressure were significantly greater during VAT pacing (VAT: 4.5 +/- 1.2 1/min, 115 +/- 28 mmHg; VOO: 3.7 +/- 0.8 1/min 105 +/- 25 mmHg respectively), although there was no difference in pulmonary artery end-diastolic pressure. Maximal exercise performance was assessed using the Bruce protocol in both pacing modes. Neither the patient nor the supervising physician was aware of the preselected pacing mode; a second physician monitored the electrocardiogram and blood pressure but influenced the point of exercise termination only if a potentially dangerous arrhythmia or hypotension occurred. Blood pressure responses were superior and atrial rates lower during VAT pacing. In all but five patients, exercise tolerance was improved by VAT pacing. This amounted to 33 percent or more in 23/35 patients. This improvement was shown to be maintained in the 20 patients who had repeat exercise tests several weeks later. Ventricular arrhythmias, hypotension, and lightheadedness frequently complicated exercise during asynchronous pacing but occurred rarely with atrial synchronized pacing. Resting hemodynamic indices did not predict the extent of improvement gained by physiological pacing.Keywords
This publication has 15 references indexed in Scilit:
- Comparison of physical work capacity and systolic time intervals with ventricular inhibited and atrial synchronous ventricular inhibited pacing.Heart, 1981
- Chronic Ventricular Pacing with Ventriculoatrial Conduction Versus Atrial Pacing in Three Patients with Symptomatic Sinus BradycardiaPacing and Clinical Electrophysiology, 1981
- Electrophysiological and haemodynamic basis for application of new pacemaker technology in sick sinus syndrome and atrioventricular block.Heart, 1979
- Three cases of hypotension and syncope with ventricular pacing: Possible role of atrial reflexesThe American Journal of Cardiology, 1978
- Hemodynamic Consequences of Atrioventricular and Ventriculoatrial PacingPacing and Clinical Electrophysiology, 1978
- Hemodynamic benefits of atrioventricular sequential pacing after cardiac surgeryThe American Journal of Cardiology, 1977
- HAEMODYNAMIC EFFECT OF ATRIAL TRIGGERED VERSUS FIXED RATE PACING AT REST AND DURING EXERCISE IN COMPLETE HEART BLOCKActa Medica Scandinavica, 1975
- Exercise stress testing in evaluation of patients with ischemic heart diseaseProgress in Cardiovascular Diseases, 1969
- Hemodynamic sequelae of atrial, ventricular, and sequential atrioventricular pacing in cardiac patientsAmerican Heart Journal, 1966
- SYNCHRONICITY OF VENTRICULAR CONTRACTION: OBSERVATIONS COMPARING HAeMODYNAMIC EFFECTS OF ATRIAL AND VENTRICULAR PACINGHeart, 1963