Effect of Gaseous Impurities on BF3 Proportional Counters

Abstract
The effect which SiF4 , SO2 , and SF6 have on the plateau and the pulse size distribution of a proportional counter were investigated. SiF4 was tested at three pressures, 30, 45, and 60 cm, of BF3 and its effect was found to be independent of the counter pressure for the range of value considered. From the variation in the plateau, the attachment probability for SiF4 was calculated to be h = 1.485×10−5 and its cross section for attachment to be σa = 5.12×10−20cm2 . The amount of these gases permitted, without the counter being affected beyond the limits of tolerance which are set up, were found to be 0.04% for SiF4 , 0.01% for SO2 , and 2.0×10−6% for SF6 .