In Vivo Validation of Magnetic Resonance Blood Volume Flow Measurements with Limited Spatial Resolution in Small Vessels
- 1 June 1995
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 33 (6) , 778-784
- https://doi.org/10.1002/mrm.1910330606
Abstract
The accuracy of magnetic resonance phase contrast volume flow measurements in small blood vessels is expected to be smaller than in large vessels, because of partial volume effects at the vessel boundary. Accuracy was validated in the dog femoral artery, diameter 3.5 ± 0.7 mm, using an ultrasonic transit‐time flowmeter (TT). The number of pixels per vessel diameter (ND) ranged from 1.6 to 4.8. The vessel cross‐section was determined using a threshold in the magnitude image. Between the two methods the correlation coefficient was 0.95 (range 10–200 ml/min). The proportional difference (PD), (QTT – QMR/1/2(QTT + QMR), was 0.8%, showing no systematic difference between the methods. The PDs standard deviation was 27%, and 19% for flow rates above 30 ml/min. Only a significant decrease of the PDs variance was found at the highest ND values, suggesting other sources of error than partial volume effects. It is concluded that with an ND value of about 3, accurate blood volume flow rates can be determined.Keywords
This publication has 23 references indexed in Scilit:
- Analysis of systematic and random error in MR volumetric flow measurementsMagnetic Resonance in Medicine, 1993
- Magnetic Resonance Velocity Measurements in Small ArteriesInvestigative Radiology, 1992
- Accuracy of colour Doppler ultrasound velocity measurements in small vesselsJournal of Biomedical Engineering, 1991
- Interleaved Magnetic Resonance and Ultrasound by Electronic SynchronizationInvestigative Radiology, 1991
- Evaluation of Magnetic Resonance Velocimetry for Steady FlowJournal of Biomechanical Engineering, 1990
- Simultaneous monitoring of coronary blood flow and 31P NMR detected myocardial metabolitesMagnetic Resonance in Medicine, 1988
- In Vivo Validation of MR Velocity ImagingJournal of Computer Assisted Tomography, 1987
- Quantitative phase-flow MR imaging in dogs by using standard sequences: comparison with in vivo flow-meter measurementsAmerican Journal of Roentgenology, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- NMR Spin-Echo Flow MeasurementsJournal of Applied Physics, 1971