Absolute efficiency and time-response measurement of single-photon detectors
- 1 April 1994
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 33 (10) , 1844-1853
- https://doi.org/10.1364/ao.33.001844
Abstract
Using correlated photons from spontaneous parametric downconversion, we have measured both the absolute quantum efficiencies and the time responses of four single-photon detectors. Efficiencies as high as (76.4 ± 2.3)% (at 702 nm) were seen, which to our knowledge are the highest reported single-photon detection efficiencies. An auxiliary retroreflection mirror was found to increase the net detection efficiency by as much as a factor of 1.19. The narrowest time profile for coincidences between two detectors displays a peak with 300 ps FWHM. We also investigated the presence of afterpulses and the effects of saturation and varying device parameters.Keywords
This publication has 8 references indexed in Scilit:
- High-efficiency single-photon detectorsPhysical Review A, 1993
- Quantum correlated twin beamsApplied Physics B Laser and Optics, 1992
- Dispersion cancellation in a measurement of the single-photon propagation velocity in glassPhysical Review Letters, 1992
- Absolute measurement of detector quantum efficiency using parametric downconversionApplied Optics, 1987
- Detection of individual 0.4–28 μm wavelength photons via impurity-impact ionization in a solid-state photomultiplierApplied Physics Letters, 1987
- Theory of parametric frequency down conversion of lightPhysical Review A, 1985
- Use of two-photon light for absolute calibration of photoelectric detectorsSoviet Journal of Quantum Electronics, 1980
- On the avalanche initiation probability of avalanche diodes above the breakdown voltageIEEE Transactions on Electron Devices, 1973