Measuring the power of an in situ intraocular lens with the keratometer
- 1 January 1988
- journal article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 14 (1) , 64-67
- https://doi.org/10.1016/s0886-3350(88)80066-x
Abstract
A method to measure the power of an intraocular lens (IOL) in situ using an extended range keratometer and a depth measuring apparatus is described. With this method, the 1OL power was measured in ten patients with an anterior chamber IOL and in ten patients with a posterior chamber IOL and compared with the,labeled power. The total mean error (measured power minus labeled power) was found to be 0.01 diopters (D) with a standard deviation of 0.22 D. The largest error was 0.40 D.Keywords