Flow dynamic studies were performed in a mechanical pulse duplicator with Starr-Edwards ball valves. It was shown that the "effective orifice" of the ball valve is less than the measured orifice of the cage of the valve. The orifice is decreased to 83 per cent of the measured orifice in the smaller and to 89 per cent in the larger valves. These measurements corresponded well with the measurements of the "effective orifice" of the valve implanted in patients who were examined by left and right heart catheterization and left ventricular angiocardiograms several months after placement of the valve. The closing regurgitation found in early systole in the mechanical system appears to be negligible in patients. Left ventricular angiograms did not show reflux of dye into the left atrium through the valve cage. The size of the mitral annulus in patients was measured from left ventricular angiograms. The size of the mitral ball valve prosthesis which the mitral annulus will accommodate could be predicted with good accuracy from these measurements.Keywords
This publication has 8 references indexed in Scilit:
- Hydraulic formula for calculation of the area of the stenotic mitral valve, other cardiac valves, and central circulatory shunts. IPublished by Elsevier ,2004
- Mitral replacement: Late results with a ball valve prosthesisProgress in Cardiovascular Diseases, 1962
- Contrast radiography in mitral regurgitationProgress in Cardiovascular Diseases, 1962
- MITRAL REPLACEMENT: THE SHIELDED BALL VALVE PROSTHESISThe Journal of Thoracic and Cardiovascular Surgery, 1961
- OPERATION OF THE ROLLER PUMP FOR EXTRACORPOREAL CIRCULATIONThe Journal of Thoracic and Cardiovascular Surgery, 1960
- Studies of the Mitral Valve. I. Anatomic Features of the Normal Mitral Valve and Associated StructuresCirculation, 1952