Design of a Load Cell for the Wagner Distractor
- 1 June 1989
- journal article
- review article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 203 (2) , 91-96
- https://doi.org/10.1243/pime_proc_1989_203_016_01
Abstract
A leg length inequality can be treated by using a Wagner distractor. With this apparatus, it is possible to extend the upper or lower leg slowly. By analysing the measurements of the axial force acting on the femur or tibia the treatment can be optimized and soft tissue damage prevented. To measure the axial force, a load cell has been designed and constructed. This load cell is built into a Wagner distractor and connected with a specially designed measuring amplifier. The load cell and amplifier do not interfere with the lengthening procedure. They are safe for the patient and have an acceptable error of 7 per cent.Keywords
This publication has 1 reference indexed in Scilit: