Cryogenic limb array etalon spectrometer calibration
- 1 November 1990
- proceedings article
- Published by SPIE-Intl Soc Optical Eng
- p. 194-202
- https://doi.org/10.1117/12.23046
Abstract
The CLAES is calibrated with a full-aperture blackbody on the instrument-aperture door. In laboratory calibration, the blackbody is resistively heated. On orbit, the blackbody is intended to be heated by exposure to radiation from the earth while the door is open; calibration data are then taken at several temperatures after closing the door, as the blackbody cools to the temperature of the instrument's cryogenic telescope. An analysis of radiometric calibration-source accuracy is shown, indicating a nominal value of 2.7 percent at 12.63 microns. Preliminary analysis of calibration data indicates a measurement repeatability of about 1.25 percent. Details of the blackbody design, construction, and thermal instrumentation are given.Keywords
This publication has 0 references indexed in Scilit: