Estimation of Imprecision in Immunoassay Quality Assessment Programmes
- 1 January 1987
- journal article
- research article
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 24 (1) , 98-106
- https://doi.org/10.1177/000456328702400115
Abstract
A computer was used to produce and store a large body of simulated thyroxine and thyrotropin radioimmunoassay results over wide concentration ranges. Sets of 24 results, designed to represent results returned to an external quality assessment survey, were randomly drawn from the stored results in order to compare limited data survey estimates of imprecision with the ‘true’ values obtained from the population data. Results show that estimation of an imprecision profile has advantages over the usual index of imprecision, but that neither form of estimate is particularly reliable when the available data is limited. Results suggest that survey organisers should place considerably less emphasis on results from any one survey and greater emphasis on analysing cumulative data and method comparison data.This publication has 8 references indexed in Scilit:
- Estimation of the response-error relationship in immunoassay.Clinical Chemistry, 1985
- A reliable method of estimating the variance function in immunoassayComputational Statistics & Data Analysis, 1985
- The Evaluation of Imprecision in Collaborative Immunoassay Quality-Assessment ProgrammesAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1984
- Estimation of a Variance Function, with Application to ImmunoassayJournal of the Royal Statistical Society Series C: Applied Statistics, 1981
- Variance function estimation for immunoassaysComputer Programs in Biomedicine, 1980
- External Quality Assessment of Serum Alpha-Fetoprotein AssaysAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1980
- A design for interlaboratory quality-control programsClinical Biochemistry, 1977
- The Form and Estimation of a Variance Function, with Particular Reference to RadioimmunoassayJournal of the Royal Statistical Society Series C: Applied Statistics, 1977