Fluorescent erythrocyte ghosts as standards for quantitative flow cytometry
Open Access
- 1 May 1995
- Vol. 20 (1) , 14-18
- https://doi.org/10.1002/cyto.990200104
Abstract
We report here a quick and inexpensive method for preparing standards of known fluorochrome content for calibration and quantitation of flow cytometry fluorescence signals. Erythrocyte ghosts prepared by hypotonic lysis are filled with solutions containing fluorescently labeled dextran. Standards prepared by this technique have a narrow range of fluorescence and a linear response of fluorescence to fluorochrome content up to 2 × 106 fluorochrome molecules/cell. The volume of ghost standard particles is roughly 70 femtoliters (fl)/cell. The fluorescence of ghost standards is nearly identical to that of commercially available microbead standards of similar fluorochrome content. Ghost standards have stable fluorescence for at least 3 weeks at 4°C. These standards can be made with any fluorochrome or combination of fluorochromes over a wide concentration range.Keywords
This publication has 9 references indexed in Scilit:
- Real-time analysis of the assembly of ligand, receptor, and G protein by quantitative fluorescence flow cytometryBiochemistry, 1991
- Interlaboratory study of cellular fluorescence intensity measurements with fluorescein‐labeled microbead standardsCytometry, 1991
- Use of Quality Control Standards in Clinical Flow CytometryAnnals of Medicine, 1991
- Mechanism of fluorescence concentration quenching of carboxyfluorescein in liposomes: Energy transfer to nonfluorescent dimersAnalytical Biochemistry, 1988
- MECHANISM OF CONCENTRATION QUENCHING OF A XANTHENE DYE ENCAPSULATED IN PHOSPHOLIPID VESICLESPhotochemistry and Photobiology, 1986
- Controls for Flow Cytometers in Hematology and Cellular ImmunologyAnnals of the New York Academy of Sciences, 1986
- A new standard fluorescence microsphere for quantitative flow cytometryJournal of Immunological Methods, 1985
- T cell subsets defined by expression of Lyt-1,2,3 and Thy-1 antigens. Two-parameter immunofluorescence and cytotoxicity analysis with monoclonal antibodies modifies current views.The Journal of Experimental Medicine, 1980
- [16] Preparation of impermeable ghosts and inside-out vesicles from human erythrocyte membranesPublished by Elsevier ,1974