Autocorrection in MR Imaging: Adaptive Motion Correction without Navigator Echoes
- 1 June 2000
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 215 (3) , 904-909
- https://doi.org/10.1148/radiology.215.3.r00jn19904
Abstract
A technique for automatic retrospective correction of motion artifacts on magnetic resonance (MR) images was developed that uses only the raw (complex) data from the MR imager and requires no knowledge of patient motion during the acquisition. The algorithm was tested on coronal images of the rotator cuff in a series of 144 patients, and the improvements in image quality were similar to those achieved with navigator echoes. The results demonstrate that autocorrection can significantly reduce motion artifacts in a technically demanding MR imaging application.Keywords
This publication has 24 references indexed in Scilit:
- MRI artifact cancellation due to rigid motion in the imaging planeIEEE Transactions on Medical Imaging, 1996
- Planar‐motion correction with use of k‐space data acquired in fourier MR imagingJournal of Magnetic Resonance Imaging, 1995
- Real‐time motion detection in spiral MRI using navigatorsMagnetic Resonance in Medicine, 1994
- Echo-planar MR imaging.Radiology, 1994
- Two-point interference method for suppression of ghost artifacts due to motionJournal of Magnetic Resonance Imaging, 1993
- Projection Reconstruction Techniques for Reduction of Motion Effects in MRIMagnetic Resonance in Medicine, 1992
- An improved algorithm for 2-D translational motion artifact correctionIEEE Transactions on Medical Imaging, 1991
- Spinal cord compression due to metastatic disease: diagnosis with MR imaging versus myelography.Radiology, 1989
- Respiratory Ordered Phase Encoding (ROPE)Journal of Computer Assisted Tomography, 1985
- Respiratory gating in magnetic resonance imaging at 0.5 Tesla.Radiology, 1984