Abstract
Superposition and calibrated attenuator methods are often used to determine the nonlinearity of photodetectors. A variation of these methods is described and applied to the calibration of optical pyrometers. This scheme allows a considerable reduction in the time needed for the measurement because only three independent sets of measurements are required, without any need to realize an accurate radiance scale on the sources.Keywords
This publication has 3 references indexed in Scilit:
- Accurate pyrometry with microsecond time resolutionJournal of Physics E: Scientific Instruments, 1979
- Accurate measurements of and corrections for nonlinearities in radiometersJournal of Research of the National Bureau of Standards Section A: Physics and Chemistry, 1972
- Numerical methods for solving linear least squares problemsNumerische Mathematik, 1965