Conceptually, transthoracic defibrillation threshold current can be considered a function of at least two quantities. It is directly proportional to the transcardiac threshold current and inversely proportional to the transcardiac current fraction (FC) or the ratio of transcardiac and transthoracic current. Although experiment and theoretical estimates of FC have been as high as 45%, previous measurements in humans have not been made. This study was designed to quantify FC in humans. During intraoperative testing of the automatic implantable cardioverter defibrillator, transthoracic rescue shocks of 200-400 J were delivered when the device failed to defibrillate. Simultaneous transthoracic voltage (VT) and transcardiac voltage (VC) between two implanted epicardial patch electrodes were measured. The ratio VC/VT, was 0.04 .+-. 0.03 (mean .+-. SD) in 10 patients. In 16 dogs, a comparison was made between direct measurement of FC and VC/VT. FC was determined with a specially designed electrode system, which was calibrated to account for field distortion introduced by the electrodes. There was no significant difference between FC and VC/VT, which were both approximately 0.05, suggesting that VC/VT was statistically equivalent to FC. The results of this study, therefore, indicate that during transthoracic defibrillation in humans, approximately 4% of transthoracic current traverses the heart. This relatively small percentage of current results from the existence of parallel pathways, such as the thoracic cage and lungs, which shunt current around the heart.This publication has 16 references indexed in Scilit:
- A volume conductor model of the thorax for the study of defibrillation fieldsIEEE Transactions on Biomedical Engineering, 1988
- Relationship between canine transthoracic impedance and defibrillation threshold. Evidence for current-based defibrillation.Journal of Clinical Investigation, 1987
- Optimal Electrode Configurations for External Cardiac Pacing and Defibrillation: An Inhomogeneous StudyIEEE Transactions on Biomedical Engineering, 1987
- The Effects of Thoracic Inhomogeneities on the Relationship Between Epicardial and Torso PotentialsIEEE Transactions on Biomedical Engineering, 1986
- The automatic implantable cardioverter-defibrillator: An overviewJournal of the American College of Cardiology, 1985
- Use of the Finite Element Method to Determine Epicardial from Body Surface Potentials Under a Realistic Torso ModelIEEE Transactions on Biomedical Engineering, 1984
- Termination of ventricular fibrillation in dogs by depolarizing a critical amount of myocardiumThe American Journal of Cardiology, 1975
- Current distribution from defibrillation electrodes in a homogeneous torso modelJournal of Electrocardiology, 1969
- Changes in the body's QRS surface potentials produced by alterations in certain compartments of the nonhomogeneous conducting modelAmerican Heart Journal, 1969
- A THROUGH TYPE CURRENT TRANSFORMER AND AMPLIFIER FOR MEASURING ALTERNATING CURRENTS OF A FEW MILLIAMPERESReview of Scientific Instruments, 1931