Silicon Junctions as Particle Spectrometers
- 1 June 1960
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IRE Transactions on Nuclear Science
- Vol. 7 (2/3) , 195-199
- https://doi.org/10.1109/tns2.1960.4315764
Abstract
Both Au-Si surface barriers and silicon p-n junctions operate satisfactorily as particle spectrometers. The use of high resistivity material gives a depletion region wide enough to stop high-energy protons, α particles, and heavier ions. A resolution of 16 kev (width at half-maximum) for 6.04 mev α particles is obtained with a junction depletion layer sufficient to stop 10 mev α particles. With the depletion region very close to the surface, ß particles in the range from 10 kev to several hundred kev are detectable. The pulse amplitude distribution from Co 57 indicates the presence of the two well-known lines at 115 and 129 kev. An attempt has been made to correlate the observed performance with the semiconductor parameters and the diode static characteristics.Keywords