Cardiac ejection fraction: Phantom study comparing cine MR imaging, radionuclide blood pool imaging, and ventriculography
- 1 March 1992
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 2 (2) , 135-142
- https://doi.org/10.1002/jmri.1880020205
Abstract
The accuracy and reproducibility of cardiac ejection fraction (EF) measurements based on cine magnetic resonance (MR) imaging, radionuclide multigated acquisition (MUGA) blood pool imaging, and angiographic ventriculography were evaluated by comparing them with a volumetrically determined standard. A biventricular, compliant, fluid‐filled heart phantom was developed to mimic normal cardiac anatomy and physiology. Ventricular EFs were measured with cine MR imaging by summation of nine contiguous 10‐mm‐thick sections in short and long axis, with single‐plane ventriculography, and with MUGA. Three measurements were performed with each modality for each of three EFs. Ventriculography was least accurate, with average relative errors ranging from 7.9% for the largest EF to 60.1% for the smallest. Cine MR was most accurate, with average relative errors ranging from 4.4% to 8.5%. MUGA EF measurements showed good correlation, with average relative errors ranging from 7.1% to 22.4%. Comparison of the error variances for the three modalities with the F test revealed that MR and MUGA EF measurements were significantly more accurate than those based on ventriculography (P <.01). No significant difference was demonstrated between the accuracy of short‐ and long‐axis cine MR acquisitions.Keywords
This publication has 42 references indexed in Scilit:
- The use of biplane angiocardiography for the measurement of left ventricular volume in manPublished by Elsevier ,2004
- Quantification of regional myocardial function by rapid cine MR imagingAmerican Journal of Roentgenology, 1988
- Magnetic resonance imaging of the heart for determination of ejection fractionInternational Journal of Cardiology, 1988
- Determination of left ventricular volumes and ejection fraction by nuclear magnetic resonance imagingAmerican Heart Journal, 1987
- FLASH imaging. Rapid NMR imaging using low flip-angle pulsesJournal of Magnetic Resonance (1969), 1986
- Quantitative contrast angiography for assessment of ventricular performance in heart diseaseJournal of the American College of Cardiology, 1983
- Quantitative imaging in nuclear cardiology.Circulation, 1982
- Real-Time Radionuclide Cineangiography in the Noninvasive Evaluation of Global and Regional Left Ventricular Function at Rest and during Exercise in Patients with Coronary-Artery DiseaseNew England Journal of Medicine, 1977
- Quantitative evaluation of left ventricular function by radiographic techniques.Circulation, 1976
- Left ventricular volume and mass from single-plane cineangiocardiogram. A comparison of anteroposterior and right anterior oblique methodsAmerican Heart Journal, 1970