Calibrating Raman Spectrometers with Plasma Lines from the Argon Ion Laser
- 1 September 1979
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 33 (5) , 475-476
- https://doi.org/10.1366/0003702794925237
Abstract
The majority of the plasma lines of the argon ion laser are shown to be transitions of the argon ion. These lines are useful for calibrating Raman spectrometers. A table is given of the plasma lines, including relative intensities, wavelengths (in standard air), wavenumber ( in vacuo), and wavenumber shifts relative to the 488.0 and 514.5 nm laser lines. Spectra showing the relative intensities of these lines are also shown.Keywords
This publication has 2 references indexed in Scilit:
- Calculation of the Raman shift in vacuumSpectrochimica Acta Part A: Molecular Spectroscopy, 1969
- Tables of Spectral Lines of Neutral and Ionized AtomsPublished by Springer Nature ,1968