Effect of silicone gel breast prosthesis on electron and photon dose distributions
- 1 January 1983
- journal article
- Published by Wiley in Medical Physics
- Vol. 10 (1) , 96-99
- https://doi.org/10.1118/1.595279
Abstract
The effect of a silicone gel breast prosthesis on the absorbed dose distribution of 9-20 MeV electron beams and 1.25-15 MV photon beams was studied. Compared to water measurements, at depths smaller than the practical range of the electron beams, the central axis depth dose values below the prosthesis were lower for all energies by as much as 3.5%. However, at depths near the practical range, the central axis depth dose values for the prosthesis were greater than that of water by as much as 33%. Since this occurs near the end of the electron range, the resultant difference may not be clinically significant. Results of the effect of breast prosthesis on photon depth dose distributions reveal that no clinically significant perturbation is produced by the breast prosthesis using Co-60, 6- and 15-MV radiations.Keywords