Dosimetric properties of a scanned beam microtron at low monitor unit settings: Importance for conformal therapy
- 1 March 1996
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 23 (3) , 329-335
- https://doi.org/10.1118/1.597662
Abstract
The dosimetric stability, linearity, dose rate dependence, and flatness of both photon and electron beams have been evaluated for a racetrack microtron at low monitor unit settings. For photons, the variation in dosimetric output about the mean is dmax and 10‐cm depths, and only slightly ≳3% at 20 cm, even at only 3 MU, in contrast with other scanned beam accelerators. Broad electron beams on the microtron are created by the superposition of the elementary beam pulses either directly from the scan magnets, or after their broadening through a scattering foil. The dosimetric instability both with and without the foil was less than 0.6% for both the 25‐ and 50‐MeV electrons. Dose nonlinearity was <1% above 10 MU. Field flatness was determined for scan matrices designed to produce a flat field both with and without a scattering foil. Symmetry and flatness deviations were <3% for both electron energies when a scattering foil was used, even for a single scan. The variation of the electron dose per monitor unit between dose rates of 85–300 MU/min was <1% (25 MeV) and <4% (50 MeV) when a scattering foil was used, but as high as 22% (25 MeV) and 36% (50 MeV) for broad beams generated by elementary beam pulses directly from the scanning magnets. The microtron exhibits dosimetric properties which fulfill the recommendations of Task Groups 21 and 25. Based on the stability of the scanned beam at low monitor unit settings, the microtron can be used for 3‐D conformal therapy with both photons and electrons.Keywords
This publication has 0 references indexed in Scilit: