A method for testing scattered radiation grids
- 1 March 1980
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 7 (2) , 165-167
- https://doi.org/10.1118/1.594666
Abstract
A simple method of evaluating contrast improvement factors and other related quantities has been developed. With this method a water or lucite phantom is imaged with a lead disk on a nonscreen film using a scattered radiation grid with unknown characteristics. The image is then compared with that obtained with the same arrangement but with a calibrated scattered radiation grid. The method described does not need sophisticated instruments and can be used in any department of diagnostic radiology.Keywords