Minimum Numbers of Fresh Whole Blood and Plasma Samples From Patients and Healthy Subjects for ISI Calibration of CoaguChek and RapidPointCoag Monitors
Open Access
- 1 June 2002
- journal article
- Published by Oxford University Press (OUP) in American Journal of Clinical Pathology
- Vol. 117 (6) , 892-899
- https://doi.org/10.1309/m1fe-ddvx-2v81-k5mp
Abstract
The international sensitivity index (ISI) calibration of point-of-care-test (POCT) prothrombin time (PT) whole blood monitors is complex, requiring manual PT testing of 60 patients’ and 20 healthy subjects’ plasma samples. The possibility of reducing these numbers was studied by a Monte Carlo Bootstrap study for 2 POCT PT systems. For reduced sample numbers, this consisted of 50,000 calibrations using whole blood and plasma samples tested on the monitors with manual PT testing of plasma samples from the same blood donations. There was little effect on mean ISI by reduction of sample numbers to a total of 7, but there was progressively less certainty regarding the reliability of the calibration. Precision of the calibrations and international normalized ratio deviation were not affected markedly by reducing numbers to half. As ISI calibration with the 2 POCT systems was less precise than conventional manual testing, for maximum confidence, reduction of numbers is not advised.Keywords
This publication has 6 references indexed in Scilit:
- European Concerted Action on Anticoagulation (ECAA): multicentre international sensitivity index calibration of two types of point‐of‐care prothrombin time monitor systemsBritish Journal of Haematology, 2002
- The effect of sample size on fresh plasma thromboplastin ISI determinationBritish Journal of Haematology, 1999
- Minimum Lyophilized Plasma Requirement for ISI CalibrationAmerican Journal of Clinical Pathology, 1998
- International Collaborative Study for the Calibration of a Proposed Reference Preparation for Thromboplastin, Human Recombinant, PlainThrombosis and Haemostasis, 1998
- Determination of the International Sensitivity Index of a New Near-Patient Testing Device to Monitor Oral Anticoagulant TherapyThrombosis and Haemostasis, 1997
- European Concerted Action On Anticoagulation (Ecaa)Pathophysiology of Haemostasis and Thrombosis, 1996