Determination of absorbed dose to water with ionisation chambers calibrated in free air for medium-energy X-rays
- 1 October 1988
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 33 (10) , 1171-1185
- https://doi.org/10.1088/0031-9155/33/10/006
Abstract
In order to determine absorbed dose to water from an in-phantom measurement by an ionisation chamber, three corrections need to be applied: (a) the ratio of water and air mass energy absorption coefficients, (b) a quality factor which corrects for the effect on the chamber response of the different photon spectra in-phantom and in-air, and (c) a perturbation factor which corrects for the water displaced by the chamber. Values of these corrections have been derived theoretically for a 0.6 cm3 cylindrical chamber for a range of medium-energy X-rays. The perturbation corrections are found to be less than 1%; these values, which are verified experimentally, agree with ICRU report 23 and not with the recent (1987) IAEA code of practice. This conclusion is expected to hold also for other small ionisation chambers. Tissue-air ratios have been derived from corrected ionisation chamber measurements and demonstrate that, for accurate dosimetry, medium-energy X-ray beams should be specified by peak tube potential as well as by half-value layer.Keywords
This publication has 7 references indexed in Scilit:
- Photon mass attenuation and energy-absorption coefficientsPublished by Elsevier ,2002
- Response of coaxial Ge(Li) detectors to narrow beams of photons for stripping of X-ray bremsstrahlung spectraNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1987
- Conversion factor f for X-ray beam qualities, specified by peak tube potential and HVL valuePhysics in Medicine & Biology, 1987
- Water calorimetric determination of absorbed dose by 280 kVp orthovoltage X-raysRadiotherapy and Oncology, 1985
- ZUR MESSUNG DER STANDARDIONENDOSIS IN FESTEN UND FLUSSIGEN MEDIEN BEI RONTGEN UND GAMMASTRAHLUNG MIT ENERGIEN ZWISCHEN 50 KEV UND 3 MEV .I. FEHLER BEI DER MESSUNG DER STANDARDIONENDOSIS IM PHANTOM MIT VERSCHIEDENEN KOMMERZIELL GEFERTIGTEN IONISATIONSKAMMERN1965