Measurement and prediction of drug solvent losses by evaporation from a jet nebuliser
- 1 May 1992
- journal article
- Published by IOP Publishing in Clinical Physics and Physiological Measurement
- Vol. 13 (2) , 129-134
- https://doi.org/10.1088/0143-0815/13/2/004
Abstract
The evaporative losses of solvent (water) from a commercial jet nebuliser (Unineb, Unimed Ltd) were estimated and found to be approximately 14% of total nebuliser output. A simple analysis of the nebulisation process then enabled the amount of the drug colomycin released as potentially respirable aerosol to be predicted from measurements of the total nebulisation time. Good agreement was found between the predicted (P) and measured (M) values such that P = 1.07 M (r2 = 0.98). The analysis also indicates that the proportion of the drug released as aerosol will increase as nebulisation proceeds provided nebuliser output is continuous and that for a given initial mass of drug a larger amount of drug is released as an aerosol if the volume of administration used in the nebuliser is as large as possible.Keywords
This publication has 6 references indexed in Scilit:
- A mathematical model of some aspects of jet nebuliser performanceClinical Physics and Physiological Measurement, 1991
- Some factors influencing the efficiency of a jet nebuliser systemClinical Physics and Physiological Measurement, 1990
- Importance of Evaporative Water Losses During Standardized Nebulized Inhalation Provocation TestsChest, 1989
- Why nebulise for more than five minutes?Archives of Disease in Childhood, 1989
- Deposition of carbenicillin aerosols in cystic fibrosis: effects of nebuliser system and breathing pattern.Thorax, 1988
- NEBULISED COLOMYCIN FOR EARLY PSEUDOMONAS COLONISATION IN CYSTIC FIBROSISThe Lancet, 1985