Measurement of brain tissue oxygenation performed using positron emission tomography scanning to validate a novel monitoring method

Abstract
Object. The benefits of measuring cerebral oxygenation in patients with brain injury are well accepted; however, jugular bulb oximetry, which is currently the most popular monitoring technique used has several shortcomings. The goal of this study was to validate the use of a new multiparameter sensor that measures brain tissue oxygenation and metabolism (Neurotrend) by comparing it with positron emission tomography (PET) scanning. Methods. A Neurotrend sensor was inserted into the frontal region of the brain in 19 patients admitted to the neurointensive care unit. After a period of stabilization, the patients were transferred to the PET scanner suite where C15O, 15O2, and H215O PET scans were obtained to facilitate calculation of regional cerebral blood volume, O2 metabolism, blood flow, and O2 extraction fraction (OEF). Patients were given hyperventilation therapy to decrease arterial CO2 by approximately 1 kPa (7.5 mm Hg) and the same sequence of PET scans was repeated. For each scanning sequence, end-c...