Abstract
Due to their high mobility, electrons represent the major component in the initial phase of the current in an ionisation chamber irradiated with short pulses of ionising radiation. The authors have developed 'fast' ionisation chambers and, by means of these, have directly measured the electron component of the chamber current. The actual number of electrons collected appeared to be significantly greater than that calculated when using attachment coefficient and drift velocity values found in the literature. For example, approximately four times the expected value is collected in room air. The deficit of negative ions resulting from the collection of electrons is incorporated within a modified version of Boag's recombination equation. The recombination loss up to an exposure per pulse of, at least, 2.58*10-2C kg-1(100R) can be calculated in this manner.Keywords
This publication has 5 references indexed in Scilit:
- Current collection and ionic recombination in small cylindrical ionization chambers exposed to pulsed radiationThe British Journal of Radiology, 1980
- Charge collection characteristics of an air ion chamber for 0.2μs pulses of 25 MeV electronsPhysics in Medicine & Biology, 1974
- Recombination in ionization chambers irradiated with pulsed electron beams. I. Plane parallel plate chamberPhysics in Medicine & Biology, 1969
- The saturation curve at high ionization intensityBritish Journal of Applied Physics, 1952
- Ionization Measurements at Very High Intensities—Part IThe British Journal of Radiology, 1950