Rational use of the PFA-100 device for screening of platelet function disorders and von Willebrand disease
- 1 June 2002
- journal article
- research article
- Published by Wolters Kluwer Health in Blood Coagulation & Fibrinolysis
- Vol. 13 (4) , 349-353
- https://doi.org/10.1097/00001721-200206000-00011
Abstract
Two hundred and five patients referred for evaluation of platelet functions and 126 healthy controls were tested with the PFA-100 instrument. A cut-off value of 150 s for collagen/epinephrine (CEPI) closure time (CT) produced most acceptable sensitivity (90%), specificity (85.2%), and positive (82.6%) and negative (91.6%) predictivity values for screening of platelet function disorders and von Willebrand disease (vWD). All patients with vWD and Glanzmann thrombasthenia could be detected by PFA-100. Both CEPI and collagen/adenosine diphosphate (CADP) CTs were elevated in all of these cases. Sensitivity of the device was 81.6% for patients with platelet secretion defects. CADP CT was normal in 63.9% of the patients in this subgroup. Specificity (47%) and positive predictivity (57%) of the instrument were diminished in patients with low hemoglobin concentrations. Depending on the results, an algorithm was developed for screening of platelet function disorders and vWD with PFA-100.Keywords
This publication has 12 references indexed in Scilit:
- Utility of the PFA‐100® for assessing bleeding disorders and monitoring therapy: a review of analytical variables, benefits and limitationsHaemophilia, 2001
- Influence of ABO blood groups on primary hemostasisTransfusion, 2001
- Assessment of Primary Hemostasis by PFA-100® Analysis in a Tertiary Care CenterThrombosis and Haemostasis, 2000
- Comparison of PFA-100 Closure Time and Template Bleeding Time of Patients with Inherited Disorders Causing Defective Platelet FunctionThrombosis Research, 1999
- Evaluation of Platelet Function with the PFA-100 System in Patients with Congenital Defects of Platelet SecretionThrombosis Research, 1999
- Use of a novel platelet function analyzer (PFA-100?) with high sensitivity to disturbances in von willebrand factor to screen for von willebrand's disease and other disordersAmerican Journal of Hematology, 1999
- Genetic Risk Factors for Superficial Vein ThrombosisThrombosis and Haemostasis, 1999
- PFA-100™ System: A New Method for Assessment of Platelet DysfunctionSeminars in Thrombosis and Hemostasis, 1998
- Relative value of diagnostic studies for von Willebrand diseaseThe Journal of Pediatrics, 1992
- The bleeding time does not predict surgical bleeding [see comments]Blood, 1991