Conformal intensity-modulated radiotherapy (IMRT) delivered by robotic linac - testing IMRT to the limit?
- 1 January 1999
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 44 (7) , 1639-1654
- https://doi.org/10.1088/0031-9155/44/7/305
Abstract
In this paper it is proposed that intensity-modulated radiotherapy (IMRT) could be delivered optimally by a short-length linac mounted on a robotic arm. The robot would allow the linac to 'plant' narrow pencils of photon radiation with any orientation (excluding zones within which the linac and couch might collide) relative to the planning target volume (PTV). The treatment is specified by the trajectory of the robot and by the number of monitor units (MUs) delivered at each robotic orientation. An inverse-planning method to determine the optimum robotic trajectory is presented. It is shown that for complex PTVs, specifically those with concavities in their outline, the conformality of the treatment is improved by the use of a complex trajectory in comparison with a less complex constrained trajectory and this improvement is quantified. It is concluded that robotic linac delivery would lead to a great flexibility in those IMRT treatments requiring very complicated dose distributions with complex 3D shapes. However, even using very fast computers, the goal of determining whether robotic linac delivery is the ultimate IMRT cannot be conclusively reached at present.Keywords
This publication has 8 references indexed in Scilit:
- CorrespondenceInternational Journal of Radiation Oncology*Biology*Physics, 1999
- The physics of radiation treatmentPhysics World, 1998
- Inverse planning with constraints to generate smoothed intensity-modulated beamsPhysics in Medicine & Biology, 1998
- Treatment planning optimization by quasi-Newton and simulated annealing methods for gamma unit treatment systemPhysics in Medicine & Biology, 1998
- Tabu search: An automatic treatment planning algorithm for conformal radiotherapyMedical Physics, 1998
- Optimization of intensity modulated beams with volume constraints using two methods: Cost function minimization and projections onto convex setsMedical Physics, 1998
- Intensity-modulated radiotherapy by means of static tomotherapy: A planning and verification studyMedical Physics, 1997
- Use of a multileaf collimator as a dynamic missing‐tissue compensatorMedical Physics, 1996