Isotope ratios and abundance sensitivity obtained with an inductively coupled plasma-time-of-flight mass spectrometer
- 1 October 1995
- journal article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 6 (10) , 920-927
- https://doi.org/10.1016/1044-0305(95)00484-u
Abstract
Isotope ratios and abundance sensitivities have been determined with an inductively coupled plasma-time-of-flight mass spectrometer (ICP-TOFMS). Abundance sensitivities are at least in the 106 range for low abundance ions that precede high abundance ions. Three methods of detection for isotope-ratio measurement have been compared. The three systems involve gated detection followed by analog integration, analog averaging, or ion counting. Gated ion counting offers excellent precision—between 0. 64 and 1. 00% relative standard deviation (RSD). These values approach those predicted from counting statistics and are comparable to those reported for other inductively coupled plasma-mass spectrometry (ICP-MS) instruments. In addition, a greater number of accumulated counts or longer analysis times would afford precisions of 0. 1% with stable gating electronics. The accuracy of the counting method is in the 1–10% range if no correction for mass bias is performed. However, this ion counting method suffers from a limited dynamic range due to pulse pileup. Constant-fraction discrimination gated integration and commercial boxcar averager techniques offer a broader dynamic range because of their analog nature, but the attainable RSD values are limited by drift in the detection systems and by the methods employed to calculate an accurate ratio. Overall, mass bias in the ICP-TOFMS is more severe than previous work in ICP-MS due primarily to detection system bias.Keywords
This publication has 17 references indexed in Scilit:
- Occurrence and reduction of noise in inductively coupled plasma mass spectrometry for enhanced precision in isotope ratio measurementJournal of Analytical Atomic Spectrometry, 1994
- Determination of 206/207Pb isotope ratios by ICP-MS in particulate matter from the north sea environmentAnalytical and Bioanalytical Chemistry, 1993
- Isotope ratio measurement of lead, neodymium and neodymium–samarium mixtures, Hafnium and Hafnium–Lutetium mixtures with a double focusing multiple collector inductively coupled plasma mass spectrometerJournal of Analytical Atomic Spectrometry, 1993
- Communication. Isotope ratio measurement by inductively coupled plasma multiple collector mass spectrometry incorporating a high efficiency nebulization systemJournal of Analytical Atomic Spectrometry, 1993
- Communication. Isotopic ratio measurement using a double focusing magnetic sector mass analyser with an inductively coupled plasma as an ion sourceJournal of Analytical Atomic Spectrometry, 1992
- Assessment of overall accuracy of lead isotope ratios determined by inductively coupled plasma mass spectrometry using batch quality control and the Youden two-sample methodJournal of Analytical Atomic Spectrometry, 1992
- Optimization of the mass scanning rate for the determination of lead isotope ratios using an inductively coupled plasma mass spectrometerJournal of Analytical Atomic Spectrometry, 1991
- Fast-Pulse Detection for Isotopic Abundance Determination By Resonance Ionization, Time-Of-Flight Mass SpectrometryInstrumentation Science & Technology, 1988
- Determination of lead isotope ratios by inductively coupled plasma-mass spectrometry (ICP-MS)Spectrochimica Acta Part B: Atomic Spectroscopy, 1987
- Isotopic ratio measurements with an inductively coupled plasma source mass spectrometerSpectrochimica Acta Part B: Atomic Spectroscopy, 1987