The Effects of Conformity and Load in Total Knee Replacement
- 1 June 2000
- journal article
- section ii
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 375 (375) , 302-312
- https://doi.org/10.1097/00003086-200006000-00036
Abstract
The effects of different conformity ratios and loads on the ultrahigh molecular weight polyethylene stress levels acting on knee implants were examined using a nonlinear, finite element analysis. The contact condition between a rigid cylinder with a radius of 30 mm and a polyethylene plate was modeled. Nonlinear behavior of polyethylene was assumed. The polyethylene plate was constructed with varying radii, with a minimal thickness of 6 mm and with a width of 40 mm. The ratio of the cylinder radius to the radius of the polyethylene plate was defined as the conformity ratio; a conformity ratio of 0 represented a flat tibial inlay, whereas the highest ratio modeled of 0.99 was nearly conforming. The conformity ratios modeled were 0, 0.2, 0.4, 0.6, 0.7, 0.8, 0.9, 0.95, and 0.99. The loads applied were 1000 N, 2000 N, 3000 N, 4000 N, 5000 N, and 6000 N. The effects of different conformity ratios and loads on the contact area (mm2), the compressive surface stress (MPa), the shear stress (MPa), and the von Mises stress (MPa) were investigated. It was found that all of these parameters were affected by changes to the conformity ratio and to a lesser extent by load changes. That is, increasing the load from 3000 N to 6000 N resulted in a surface and shear stress increase lower than the increase in stress caused by the small change of the conformity ratio from 0.99 to 0.95. The effect of an increasing conformity ratio on the reduction in stress was more pronounced for conformity ratios above 0.8. In addition, the effect of a load increase for a flat tibial inlay was two times greater than for one with near full conformity.Keywords
This publication has 21 references indexed in Scilit:
- Finite element analysis of a novel design approach to resisting total hip dislocationClinical Biomechanics, 1999
- NUMBER OF POLYETHYLENE PARTICLES AND OSTEOLYSIS IN TOTAL JOINT REPLACEMENTSThe Journal of Bone and Joint Surgery. British volume, 1997
- JOINT LOAD CONSIDERATIONS IN TOTAL KNEE REPLACEMENTThe Journal of Bone and Joint Surgery. British volume, 1997
- Characterization of Debris Adjacent to Failed Knee Implants of 3 Different DesignsClinical Orthopaedics and Related Research, 1996
- The redundant nature of locomotor optimization lawsJournal of Biomechanics, 1995
- Optimization of the bearing surface geometry of total kneesJournal of Biomechanics, 1994
- The histological effects of the implantation of different sizes of polyethylene particles in the rabbit tibiaJournal of Biomedical Materials Research, 1990
- Bearing Surface Design in Total Knee ReplacementEngineering in Medicine, 1988
- Walking activities and wear of prostheses.Annals of the Rheumatic Diseases, 1985
- Moments of force and mechanical power in joggingJournal of Biomechanics, 1983