Simple nonperturbing temperature probe for microwave/radio frequency dosimetry
- 1 January 1989
- journal article
- research article
- Published by Wiley in Bioelectromagnetics
- Vol. 10 (2) , 209-213
- https://doi.org/10.1002/bem.2250100210
Abstract
We present a simple readout device that fills the void produced by the gradual disappearance of the Vitek model 101 Electrothermia Monitor. The new device uses commercially available probes that are similar to Bowman's (IEEE Trans Microwave Theory Tech MTT‐24:43–45, 1976) original design. As described, the device covers the range of 7 ° to 45 °C with an accuracy and resolution of better than 0.1 °C throughout. The digital readout (3½ digits) is proportional to the thermistor resistance in the tip of the probe; it is converted to temperature through a formula or printed table. Outdoor dosimetric comparisons between the new and Vitek devices were conducted. Results showed no significant difference in the calculated specific absorption rates (SARs); moreover, variance was lower in the data collected from the new device.Keywords
This publication has 4 references indexed in Scilit:
- The nonmetallic thermocouple: A differential‐temperature probe for use in microwave fieldsRadio Science, 1979
- Thermometry considerations in localized hyperthermiaMedical Physics, 1978
- A Probe for Measuring Temperature in Radio-Frequency-Heated Material (Short Papers)IEEE Transactions on Microwave Theory and Techniques, 1976
- A Nonperturbing Temperature Sensor for Measurements in Electromagnetic FieldsJournal of Microwave Power, 1974