High-field-strength MR imaging and metallic biomedical implants: an ex vivo evaluation of deflection forces
- 1 August 1988
- journal article
- research article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 151 (2) , 389-392
- https://doi.org/10.2214/ajr.151.2.389
Abstract
Ferromagnetic biomedical implants are considered a contraindication for MR imaging primarily because of the potential hazards associated with their movement or dislodgment. Many metallic biomedical implants are composed of nonferromagnetic materials and do not present a danger to patients during MR imaging. Therefore, to evaluate the ferromagnetic qualities of 36 different metallic biomedical implants (four aneurysm clips, six hemostatic clips, four dental implants, seven prosthetic heart valves, eight orthopedic prostheses, one artificial urinary sphincter, three contraceptive diaphragms, and three cerebral ventricular shunt tube connectors) not previously evaluated with a high-field-strength MR system, we measured deflection forces at the portal of the magnet of a 1.5-T MR system. Fourteen of the 36 metallic biomedical implants were determined to be ferromagnetic as indicated by their deflection in the static magnetic field. However, only the four aneurysm clips (Drake, Mayfield, McFadden, and Sundt-Kee...This publication has 8 references indexed in Scilit:
- Dental material artifacts on MR images.Radiology, 1988
- MR imaging artifacts, ferromagnetism, and magnetic torque of intravascular filters, stents, and coils.Radiology, 1988
- MR imaging in patients with metallic implants.Radiology, 1985
- The effect on medical metal implants by magnetic fields of magnetic resonance imagingSkeletal Radiology, 1985
- Magnetic resonance imaging of prosthetic heart valves.Radiology, 1985
- Magnetic resonance imaging of postoperative patients with metallic implantsAmerican Journal of Roentgenology, 1984
- Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging.Radiology, 1983
- Potential hazards in NMR imaging: heating effects of changing magnetic fields and RF fields on small metallic implantsAmerican Journal of Roentgenology, 1981