The effect of femoral head diameter upon lubrication and wear of metal-on-metal total hip replacements
- 1 February 2001
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 215 (2) , 161-170
- https://doi.org/10.1243/0954411011533724
Abstract
It has been found that a remarkable reduction in the wear of metal-on-metal hip joints can be achieved by simply increasing the diameter of the joint. A tribological evaluation of metal-on-metal joints of 16, 22,225, 28 and 36 mm diameter was conducted in 25 per cent bovine serum using a hip joint simulator. The joints were subject to dynamic motion and loading cycles simulating walking for both lubrication and wear studies. For each size of joint in the lubrication study, an electrical resistivity technique was used to detect the extent of surface separation through a complete walking cycle. Wear of each size of joint was measured gravimetrically in wear tests of at least 2 × 106 cycles duration. Joints of 16 and 22.225mm diameter showed no surface separation in the lubrication study. This suggested that wear would be proportional to the sliding distance and hence joint size in this boundary lubrication regime. A 28 mm diameter joint showed only limited evidence of surface separation suggesting that these joints were operating in a mixed lubrication regime. A 36 mm diameter joint showed surface separation for considerable parts of each walking cycle and hence evidence of the formation of a protective lubricating film. Wear testing of 16 and 22.225mm diameter metal-on-metal joints gave mean wear rates of 4.85 and 6.30mm3/106 cycles respectively. The ratio of these wear rates, 0.77, is approximately the same as the joint diameters ratio, 16/22.225 or 0.72, as expected from simple wear theory for dry or boundary lubrication conditions. No bedding-in was observed with these smaller diameter joints. For the 28 mm diameter joint, from 0 to 2 × 106 cycles, the mean wear rate was 1.62 mm3/106 cycles as the joints bedded-in. Following bedding-in, from 2.0 × 106 to 4.7 × 106 cycles, the wear rate was 0.54mm3/106 cycles. As reported previously by Goldsmith in 2000 [1], the mean steady state wear rate of the 36 mm diameter joints was lower than those of all the other diameters at 0.07 mm3/106 cycles. For a range of joints of various diameters, subjected to identical test conditions, mean wear rates differed by almost two orders of magnitude. This study has demonstrated that the application of sound tribological principles to prosthetic design can reduce the wear of metal-on-metal joints, using currently available materials, to a negligible level.Keywords
This publication has 36 references indexed in Scilit:
- A tribological study of UHMWPE acetabular cups and polyurethane compliant layer acetabular cupsJournal of Biomedical Materials Research, 2000
- Long-term registration has improved the quality of hip replacement: A review of the Swedish THR Register comparing 160,000 casesActa Orthopaedica, 2000
- Metal on Metal Total Hip Replacement Workshop Consensus DocumentClinical Orthopaedics and Related Research, 1996
- Hydrodynamic Lubrication Analysis of Metallic Hip JointTribology Transactions, 1996
- Wear mechanisms of UHMWPE in total joint replacementsWear, 1995
- Changes Seen in Lymph Nodes Draining the Sites of Large Joint ProsthesesThe American Journal of Surgical Pathology, 1989
- The use of dense alumina–alumina ceramic combination in total hip replacementJournal of Biomedical Materials Research, 1988
- Laboratory wear tests and clinical observations of the penetration of femoral heads into acetabular cups in total replacement hip joints: III: The measurement of internal volume changes in explanted Charnley sockets after 2–16 years in vivo and the determination of wear factorsWear, 1985
- Elastohydrodynamic Lubrication of Elliptical Contacts for Materials of Low Elastic Modulus I—Fully Flooded ConjunctionJournal of Lubrication Technology, 1978
- The tribology (friction, lubrication and wear) of all-metal artificial hip jointsWear, 1971