Comparison of Pulsed Source with Continuous Wave Source Excitation in Atomic and Molecular Luminescence Spectrometry via Signal/Noise Ratio Calculations
- 1 July 1977
- journal article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 31 (4) , 307-317
- https://doi.org/10.1366/000370277774463634
Abstract
Pulsed source excitation-gated detection (with or without time resolution) and cw excitation-cw detection are compared theoretically with respect to signal/noise (S/N) ratios in atomic and molecular luminescence spectrometry. The calculations indicate that in those cases where there is no loss in flux due to pulsing, pulsed source excitation with delayed gated detection results in a considerable improvement in S/N as long as the major noise sources are either not related to the source decay or, if related to the source decay or luminescence decay of impurities, the delay time between excitation and detection is greater than the corresponding source or impurity decay times. In general, pulsed source excitation-gated detection and modulated source lock-in detection do not provide significant S/N advantages over cw excitation and cw detection unless the major source of noise is related to either the detector or to non-source-induced background. Calculations of S/N ratio are given for a variety of analytically useful situations in atomic fluorescence spectrometry, molecular fluorescence spectrometry of gas phase and condensed phase species, and molecular phosphorescence spectrometry.Keywords
This publication has 41 references indexed in Scilit:
- A comparison of signal-to-noise ratios for single channel methods (sequential and multiplex) vs multichannel methods in optical spectroscopySpectrochimica Acta Part B: Atomic Spectroscopy, 1976
- Flame emission, atomic absorption, and atomic fluorescence spectrometryAnalytical Chemistry, 1976
- Fluorometric and phosphorometric analysisAnalytical Chemistry, 1976
- Review of Multielement Atomic Spectroscopic MethodsApplied Spectroscopy, 1975
- Pulsed Source Atomic Fluorescence SpectrometryApplied Spectroscopy Reviews, 1973
- Molecular Scattering of LightPublished by Springer Nature ,1968
- Derivation of Expression for Integrated Luminescence Intensity When Using Pulsing Techniques in Luminescence Spectrometry.Analytical Chemistry, 1966
- The Influence of Phosphoroscope Design on the Measured Phosphorescence Intensity in Phosphorimetry.Analytical Chemistry, 1966
- Limiting Parameters and Generalized Characteristics of Xenon LampsApplied Optics, 1963
- Design and Operation of Xenon FlashtubesJournal of Applied Physics, 1963