Durability of pyrolytic carbon‐containing heart valve prostheses
- 13 September 1982
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 16 (5) , 559-570
- https://doi.org/10.1002/jbm.820160504
Abstract
To assess abrasive wear of mechanical valve prostheses containing pyrolytic carbon components, we recovered at necropsy or surgery and analyzed by scanning electron microscopy and surface profilometry eight prostheses. Seven were implanted for 30–85 (mean 50 months). Valves included DeBakey aortic (2), DeBakey mitral (1), Beall mitral (2), Bjork–Shiley aortic (1), Cooley–Cutter mitral (1), and Lillehei–Kaster (L–K) tricuspid (1). All carbon occluders had undetectable wear. Carbon cage struts had a superficial burnish. Metallic struts had insignificant wear marks. In contrast, a Teflon Beall Model 104 valve implanted for 34 days and similarly analyzed had considerable material loss from the cage struts. This study suggests that clinically important abrasive wear will not be a late complication of cardiac valve replacement with pyrolytic carbon prostheses.This publication has 30 references indexed in Scilit:
- Aortic and Mitral Prosthetic Valve ReoperationsArchives of Surgery, 1979
- Prosthetic Aortic ValvesArchives of Surgery, 1979
- Repeat open intracardiac operationThe Journal of Thoracic and Cardiovascular Surgery, 1978
- The Determinants of Survival Following Reoperation on Prosthetic Cardiac ValvesThe Annals of Thoracic Surgery, 1978
- Replacement of prosthetic heart valvesThe Journal of Thoracic and Cardiovascular Surgery, 1977
- The pathology of wear in the Beall model 104 heart valve prosthesis.Circulation, 1977
- Cardiac pathology after valve replacement with a tilting disc prosthesis (Björk-Shiley type)The American Journal of Cardiology, 1976
- The Wear Resistance of Pure and Silicon-Alloyed Isotropic CarbonsBiomaterials, Medical Devices, and Artificial Organs, 1974
- The Behavior of Isotropic Pyrolytic Carbons under Cyclic LoadingBiomaterials, Medical Devices, and Artificial Organs, 1974
- Phonocardiographic Diagnosis of Aortic Ball VarianceCirculation, 1968