Energy and spatial dependence of the electron detection efficiencies of single channel electron multipliers used in electron spectroscopy
- 1 January 1991
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 62 (1) , 62-68
- https://doi.org/10.1063/1.1142283
Abstract
Measurements for axially incident electrons are reported for the energy and spatial dependencies of the detection efficiency of a Philips X919BL single channel electron multiplier (CEM) in both the pulse counting and analog amplification modes. The behavior may be understood in terms of the general theory reported earlier.1 Essentially there are three regions to consider: the horn, the channel and the edge of the channel. In the pulse counting mode the difference in efficiencies between the channel and horn may reach a factor of two whereas in the analog mode the difference between the channel edge and the horn may reach a factor of five. The same behavior occurs when the CEM is used as the output stage of an electron spectrometer. The above variations may lead to (i) loss of efficiency in spectrometers, (ii) spectral distortion arising from the associated presence of stray magnetic fields, and (iii) irreproducibility of spectra from different regions of the sample surface, unless specific measures, such as larger horns, off-axis channels or lens defocusing, are used to eliminate them.Keywords
This publication has 7 references indexed in Scilit:
- Channel electron multipliers: quantitative intensity measurement—efficiency, gain, linearity and bias effectsJournal of Electron Spectroscopy and Related Phenomena, 1990
- Channel electron multiplier efficiencies: the effect of the pulse height distribution on spectrum shape in auger electron spectroscopyJournal of Electron Spectroscopy and Related Phenomena, 1989
- Quantitative AES: The establishment of a standard reference spectrum for the accurate determination of spectrometer transmission functionsSurface and Interface Analysis, 1988
- Auger electron spectroscopy: Method for the accurate measurement of signal and noise and a figure of merit for the performance of AES instrument sensitivityReview of Scientific Instruments, 1988
- Quantitative aes: the problems of the energy dependent phase shift and modulation amplitude and of the non-ideal behaviour of the channel electron multiplierJournal of Electron Spectroscopy and Related Phenomena, 1987
- Secondary electron yield in the Bendix channel electron multiplierNuclear Instruments and Methods, 1977
- The Calibration of Electrostatic Analyzers and Channel Electron Multipliers Using Laboratory Simulated Omnidirectional Electron BeamsReview of Scientific Instruments, 1973