Metal-film thermopiles for use with rabbit right ventricular papillary muscles

Abstract
A method of fabricating Hill-Downing type, planar thermopiles by vacuum-deposition techniques is described in detail. The present model was designed for initial heat measurements on rabbit papillary muscles as small as 1 mg blotted wt, but it is also suitable for small bundles of frog muscle fibers (30–75). The thermopile has 20 or 14 junctions, an active length of 5 or 3.5 mm, and an actual thickness of 20 micrometer. It has an effective heat capacity of about 0.3 mcal/degrees C, a heat loss coefficient of about 0.3 mcal/degrees C - s, a temperature sensitivity of 1.4 mV/degrees C (20 junctions), and an electrical resistance of 180–200 omega. Infrared-emitting diodes are used to heat the thermopile and muscle artificially for thermal time constant and conduction-delay measurements. Performance of the thermopiles is demonstrated with initial heat records from rabbit right ventricular papillary muscles and a bundle of frog semitendinosus muscle fibers. Results of preliminary experiments concerning latency for heat generation, initial rate of heat generation, and activation heat in both types of muscles are presented.

This publication has 5 references indexed in Scilit: