Generalized reconstruction of phase contrast MRI: Analysis and correction of the effect of gradient field distortions
Open Access
- 26 September 2003
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 50 (4) , 791-801
- https://doi.org/10.1002/mrm.10582
Abstract
To characterize gradient field nonuniformity and its effect on velocity encoding in phase contrast (PC) MRI, a generalized model that describes this phenomenon and enables the accurate reconstruction of velocities is presented. In addition to considerable geometric distortions, inhomogeneous gradient fields can introduce deviations from the nominal gradient strength and orientation, and therefore spatially‐dependent first gradient moments. Resulting errors in the measured phase shifts used for velocity encoding can therefore cause significant deviations in velocity quantification. The true magnitude and direction of the underlying velocities can be recovered from the phase difference images by a generalized PC velocity reconstruction, which requires the acquisition of full three‐directional velocity information. The generalized reconstruction of velocities is applied using a matrix formalism that includes relative gradient field deviations derived from a theoretical model of local gradient field nonuniformity. In addition, an approximate solution for the correction of one‐directional velocity encoding is given. Depending on the spatial location of the velocity measurements, errors in velocity magnitude can be as high as 60%, while errors in the velocity encoding direction can be up to 45°. Results of phantom measurements demonstrate that effects of gradient field nonuniformity on PC‐MRI can be corrected with the proposed method. Magn Reson Med 50:791–801, 2003. Published 2003 Wiley‐Liss, Inc.Keywords
This publication has 16 references indexed in Scilit:
- Mr geometric distortion correction for improved frame‐based stereotaxic target localization accuracyMagnetic Resonance in Medicine, 1995
- Method for correcting magnetic resonance image distortion for frame-based stereotactic surgery, with preliminary resultsJournal of Image Guided Surgery, 1995
- Quantifying MRI geometric distortion in tissueMagnetic Resonance in Medicine, 1994
- Correction of spatial distortion in magnetic resonance angiography for radiosurgical treatment planning of cerebral arteriovenous malformationsMagnetic Resonance Imaging, 1992
- Analysis of machine-dependent and object-induced geometric distortion in 2DFT MR imagingMagnetic Resonance Imaging, 1992
- Simulation of the influence of magnetic field inhomogeneity and distortion correction in MR imagingMagnetic Resonance Imaging, 1990
- In Vivo Validation of MR Velocity ImagingJournal of Computer Assisted Tomography, 1987
- Blood Flow Imaging by Cine Magnetic ResonanceJournal of Computer Assisted Tomography, 1986
- Magnet field profiling: Analysis and correcting coil designMagnetic Resonance in Medicine, 1984
- A flow velocity zeugmatographic interlace for NMR imaging in humansMagnetic Resonance Imaging, 1982