Slit-scanning confocal microendoscope for high-resolution in vivo imaging
- 1 December 1999
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 38 (34) , 7133-7144
- https://doi.org/10.1364/ao.38.007133
Abstract
We discuss the design and construction of a novel imaging system in which a fiber-optic imaging bundle and miniature optical and mechanical components are used to allow confocal fluorescence microscopy in remote locations. The instrumentation has been developed specifically for cellular examination of tissue for optical biopsy. Miniaturization of various components makes the device usable in a clinical setting. The numerical aperture of the beam in the tissue is 0.5, and the field of view is 430 µm. The measured lateral resolution of the system is 3.0 µm. The axial point and the axial planar response functions of the confocal system were measured with a FWHM of 10 and 25 µm, respectively. In vitro and in vivo images obtained with cell cultures, human tissue specimens, and animal models indicate that the performance of the device is adequate for microscopic evaluation of cells.Keywords
This publication has 7 references indexed in Scilit:
- IntroductionOptics Express, 1998
- Silicon-micromachined scanning confocal optical microscopeJournal of Microelectromechanical Systems, 1998
- Real‐time white light reflection confocal microscopy using a fibre‐optic bundleScanning, 1997
- Imaging spectroscopy of the human ocular fundus in vivoJournal of Biomedical Optics, 1997
- Precision-cut tissue slices: Applications in pharmacology and toxicologyLife Sciences, 1995
- Confocal microscopy through a fiber-optic imaging bundleOptics Letters, 1993
- Spread Functions and Modulation Transfer Functions of Fibre-optic BundlesJournal of Modern Optics, 1988