Attenuation correction for small animal PET tomographs
- 6 April 2005
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (8) , 1837-1850
- https://doi.org/10.1088/0031-9155/50/8/014
Abstract
Attenuation correction is one of the important corrections required for quantitative positron emission tomography (PET). This work will compare the quantitative accuracy of attenuation correction using a simple global scale factor with traditional transmission-based methods acquired either with a small animal PET or a small animal x-ray computed tomography (CT) scanner. Two phantoms (one mouse-sized and one rat-sized) and two animal subjects (one mouse and one rat) were scanned in CTI Concorde Microsystem's microPET Focus for emission and transmission data and in ImTek's MicroCAT II for transmission data. PET emission image values were calibrated against a scintillation well counter. Results indicate that the scale factor method of attenuation correction places the average measured activity concentration about the expected value, without correcting for the cupping artefact from attenuation. Noise analysis in the phantom studies with the PET-based method shows that noise in the transmission data increases the noise in the corrected emission data. The CT-based method was accurate and delivered low-noise images suitable for both PET data correction and PET tracer localization.Keywords
This publication has 14 references indexed in Scilit:
- Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging.2005
- In vivomolecular and genomic imaging: new challenges for imaging physicsPhysics in Medicine & Biology, 2004
- New, faster, image-based scatter correction for 3D PETIEEE Transactions on Nuclear Science, 2000
- Attenuation correction for a combined 3D PET/CT scannerMedical Physics, 1998
- Detailed investigation of transmission and emission data smoothing protocols and their effects on emission imagesIEEE Transactions on Nuclear Science, 1996
- Singles transmission in volume-imaging PET with a137Cs sourcePhysics in Medicine & Biology, 1995
- Post injection transmission scanning in a volume imaging PET cameraIEEE Transactions on Nuclear Science, 1994
- Attenuation correction in PET using single photon transmission measurementMedical Physics, 1994
- MRI-PET Registration with Automated AlgorithmJournal of Computer Assisted Tomography, 1993
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1979