Flow-through photometric sensor for determination of sulfonamides
- 1 January 1994
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 119 (7) , 1625-1628
- https://doi.org/10.1039/an9941901625
Abstract
A sensor based on integration of spectrophotometric detection and retention of the product of the Bratton–Marshall reaction for the determination of sulfanilamide, sulfamethazine, sulfaquinoxaline and sulfathiazole is reported. The methods developed using this sensor in a flow injection manifold feature linear ranges for these analytes between 10 and 1200 ng ml–1 with relative standard deviations smaller than 2.6% in all instances and they were applied to the determination of sulfonamides in spiked milk and water samples with excellent results. The two main advantages of these sensor-based methods over other automated and non-automated counterparts are enhanced sensitivity as a result of in situ concentration and avoidance of using a de-bubbler to eliminate the nitrogen produced in the flowing system.Keywords
This publication has 15 references indexed in Scilit:
- Sensitivity Enhancement by Using an HPLC Flow-Through Sensor for Determination of Pesticide MixturesJournal of Liquid Chromatography, 1992
- Flow-injection spectrophotometric determination of arylamines and sulphonamides by diazotization and coupling in a micellar mediumAnalytica Chimica Acta, 1991
- Flow-through sensor for the direct determination of pesticide mixtures without chromatographic separationAnalytical Chemistry, 1991
- Spectrofluorimetric flow-through sensor for the determination of beryllium in alloysThe Analyst, 1991
- Enhancement of flow injection optosensing by sorbent extraction and reaction rate measurementAnalytical Chemistry, 1990
- Fluorimetric sensor for the determination of fluoride at the nanograms per millilitre levelAnalytica Chimica Acta, 1990
- Integrated retention/spectrophotometric detection in flow-injection analysisAnalytica Chimica Acta, 1989
- Automated flow-injection determination of sulfonamides by the bratton—marshall reaction for clinical analysis, assays and dissolution studies of formulationsAnalytica Chimica Acta, 1988
- An automated flow-injection serial dynamic dialysis technique for drug-protein binding studiesAnalytica Chimica Acta, 1986
- Evaluation of dual wavelength spectrophotometry for drug level monitoringAnalytical Chemistry, 1984