Practical T2 quantitation for clinical applications
- 1 September 1992
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 2 (5) , 541-553
- https://doi.org/10.1002/jmri.1880020512
Abstract
Many studies have investigated the use of magnetic resonance relaxation times for tissue characterization. A number have been performed in vivo with clinical whole-body imagers. Unfortunately, the results have yet to establish the role of quantitative tissue relaxation time measurements in clinical diagnosis. One of the major problems is that the techniques used in many of these studies are error prone, making the results inconclusive. In the present study, the problems associated with clinical T2 measurements were systematically evaluated in an attempt to obtain reliable in vivo quantitation. The authors demonstrate that spoiler gradients are the most effective technique for artifact suppression but that they render ineffective radio-frequency phase schemes such as the Meiboom-Gill modification to the Carr-Purcell sequence to compensate experimental imperfections. The present study results in a more reliable multiecho sequence for T2 measurement. Preliminary clinical results in brain and cervix demonstrate the performance of the new technique.Keywords
This publication has 48 references indexed in Scilit:
- A selective adiabatic spin-echo pulseJournal of Magnetic Resonance (1969), 1989
- Amplitude- and frequency/phase-modulated refocusing pulses that induce plane rotations even in the presence of inhomogeneous B1 fieldsJournal of Magnetic Resonance (1969), 1988
- Multiecho imaging sequences with low refocusing flip anglesJournal of Magnetic Resonance (1969), 1988
- Variable-rate selective excitationJournal of Magnetic Resonance (1969), 1988
- Iterative carr-purcell trainsJournal of Magnetic Resonance (1969), 1988
- A phase-cycling technique for canceling spurious echoes in NMR imagingJournal of Magnetic Resonance (1969), 1987
- Reproducibility of relaxation and spin-density parameters in phantoms and the human brain measured by MR imaging at 1.5TMagnetic Resonance in Medicine, 1986
- Multiple‐spin‐echo imaging with a 2D Fourier methodMagnetic Resonance in Medicine, 1986
- Compensation for pulse imperfections in NMR spin-echo experimentsJournal of Magnetic Resonance (1969), 1981
- Errors in measurements of transverse relaxation ratesJournal of Magnetic Resonance (1969), 1973