The reproducibility of single photon absorptiometry in a clinical setting
- 1 May 1989
- journal article
- research article
- Published by Springer Nature in European Journal of Nuclear Medicine and Molecular Imaging
- Vol. 15 (5) , 269-273
- https://doi.org/10.1007/bf00257547
Abstract
The reproducibility of single photon absorptiometry (SPA) results for detection of changes in bone mineral content (BMC) was evaluated in a clinical setting. During a period of 18 months with 4 different sources, the calibration scans of an aluminium standard had a variation of less than 1% unless the activity of the 125I source was low. The calibration procedure was performed weekly and this was sufficient to correct for drift of the system. The short term reproducibility in patients was assessed with 119 duplicate measurements made in direct succession. The best reproducibility (CV=1.35%) was found for fat corrected BMC results expressed in g/cm, obtained at the site proximal to the 8 mm space between the radius and ulna. Analysis of all SPA scans made during 1 year (487 scans) showed a failure of the automatic procedure to detect the space of 8 mm between the forearm bones in 19 scans (3.9%). A space adjacent to the ulnar styloid was taken as the site for the first scan in these examinations. This problem may be recognized and corrected relatively easy. A significant correlation was found between BMC of the lower arm and BMC of the lumbar spine assessed with dual photon absorptiometry. However, the error of estimation of proximal BMC (SEE=20.0%) and distal BMC (SEE=19.4%) made these measurements of little value to predict BMC at the lumbar spine in individuals. The short term reproducibility in patients combined with the long term stability of the equipment in our clinical setting showed that SPA is a reliable technique to assess changes in bone mass at the lower arm of 4% between 2 measurements with a confidence level of 95%.This publication has 15 references indexed in Scilit:
- ERRORS IN LONGITUDINAL MEASUREMENTS OF BONE-MINERAL - EFFECT OF SOURCE STRENGTH IN SINGLE AND DUAL PHOTON-ABSORPTIOMETRY1987
- USEFULNESS OF REGIONAL BONE MEASUREMENTS IN PATIENTS WITH OSTEOPOROTIC FRACTURES OF THE SPINE AND DISTAL FOREARM1987
- Involutional OsteoporosisNew England Journal of Medicine, 1986
- Should Perimenopausal Women Be Screened for Osteoporosis?Annals of Internal Medicine, 1986
- Longitudinal changes in bone mass after one year as measured by different techniques in patients with osteoporosisCalcified Tissue International, 1986
- Rates of bone loss in the appendicular and axial skeletons of women. Evidence of substantial vertebral bone loss before menopause.Journal of Clinical Investigation, 1986
- COMPARISON OF SINGLE-PHOTON AND DUAL-PHOTON ABSORPTIOMETRY IN POSTMENOPAUSAL BONE-MINERAL LOSS1985
- Assessment of bone mineral. Part 2.1984
- DOES BONE-MEASUREMENT ON THE RADIUS INDICATE SKELETAL STATUS - CONCISE COMMUNICATION1984
- Differential changes in bone mineral density of the appendicular and axial skeleton with aging: relationship to spinal osteoporosis.Journal of Clinical Investigation, 1981