Diazo-based assay for total bilirubin in a coated thin film evaluated.
Open Access
- 1 August 1984
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 30 (8) , 1310-1313
- https://doi.org/10.1093/clinchem/30.8.1310
Abstract
We compared results for total bilirubin as measured on a coated thin film and by the Evelyn-Malloy and Jendrassik-Gróf methods. We examined serum samples from patients and studied the effects of protein, hemoglobin, and lipids on bilirubin measurement. Results from the thin-film assay compared favorably with those of the other methods. Total and within-day precision (CV), assessed over a one-year period, were better than 6% and 3%, respectively, at all concentrations. Analytical recovery was 99 +/- 3%. Samples from individuals having a wide range of liver diseases demonstrated, by linear regression, good correlation between the thin-film method and the two wet-chemistry methods (correlation coefficients of 0.990 and 0.994). We conclude that the thin-film method offers a valid alternative assay for total bilirubin.This publication has 3 references indexed in Scilit:
- A diazo-based dry film for determination of total bilirubin in serum.Clinical Chemistry, 1983
- Jendrassik--Grof analysis for total and direct bilirubin in serum with a centrifugal analyzer.Clinical Chemistry, 1981
- MECHANISM OF INTERFERENCE BY HEMOGLOBIN IN THE DETERMINATION OF TOTAL BILIRUBIN .1. METHOD OF MALLOY-EVELYN1980