Signal detection in digital chest-phantom images acquired with an image intensifier
- 1 February 1989
- journal article
- digital imaging-basics
- Published by Springer Nature in Journal of Digital Imaging
- Vol. 2 (1) , 42-47
- https://doi.org/10.1007/bf03168015
Abstract
Signal detection performance was evaluated on the basis of ROC analysis using both digital and conventional images of a humanoid chest phantom. Simulated focal (coin) lesions were the target pathology. Digital images were acquired using a 57-cm image intensifier, digitized to 1024×1024×10 bits, and compared, in both video and laser-printed film formats, with conventional 14×17-inch chest films. Signal detection using digital video and laser printed images, of the same image polarity as conventional images, was found not to differ significantly from that achieved using conventional images, despite the smaller size of the digital images.Keywords
This publication has 22 references indexed in Scilit:
- Total digital radiology department: spatial resolution requirementsAmerican Journal of Roentgenology, 1987
- The total digital radiology department: an alternative viewAmerican Journal of Roentgenology, 1985
- Total digital radiology: is it feasible? Or desirable?American Journal of Roentgenology, 1984
- Statistical Approaches to the Analysis of Receiver Operating Characteristic (ROC) CurvesMedical Decision Making, 1984
- Digital radiography of the chest: clinical experience with a prototype unit.Radiology, 1983
- Evaluating imaging methodsAmerican Journal of Roentgenology, 1982
- The meaning and use of the area under a receiver operating characteristic (ROC) curve.Radiology, 1982
- Detection accuracy in chest radiographyAmerican Journal of Roentgenology, 1978
- Effect of size and position on chest lesion detectionAmerican Journal of Roentgenology, 1977
- Primer on Certain Elements of Medical Decision MakingNew England Journal of Medicine, 1975