Observer and Biological Variation of a Rapid Whole Blood D-Dimer Test
- 1 January 1998
- journal article
- review article
- Published by Georg Thieme Verlag KG in Thrombosis and Haemostasis
- Vol. 79 (01) , 91-93
- https://doi.org/10.1055/s-0037-1614226
Abstract
Summary: In consecutive patients with suspected venous thromboembolism the interobserver variability of the SimpliRED D-dimer test was evaluated by two observers who independently scored one plate, the between assay variation was performed simultaneously by a third independent observer, who assessed a second plate. The biological variation was studied, 1-4 hours later by an independent evaluation.A total of 155 patients entered the study, venous thromboembolism was present in 42 patients (28%). The interobserver variability was 2/83 samples, with a kappa of 0.95 (95% confidence interval 0.88-1.0). The between assay variation was 2/98, with a kappa value of 0.96 (95% confidence interval 0.90-1.0). When testing the biological variation the observers disagreed in 2 of 69 patients (3%).The SimpliRED D-dimer assay has a good to excellent interobserver variability, between assay variation and reproducibility.Keywords