Intramural Coronary Vasculature Prevents Transmural Radiofrequency Lesion Formation
- 8 April 2003
- journal article
- Published by Wolters Kluwer Health in Circulation
- Vol. 107 (13) , 1797-1803
- https://doi.org/10.1161/01.cir.0000058705.97823.f4
Abstract
Blood flow near a radiofrequency (RF) lesion can reduce lesion size by convective cooling. It is unknown whether blood flow through small vasculature within an RF lesion can prevent transmural lesion formation. In 40 rabbit right ventricle preparations, 2 epicardial RF lesions were created straddling a selectively perfused (0 to 12 mL/min) marginal artery (diameter, 0.34+/-0.1 mm). RF lesions were created at either 60 degrees C or 80 degrees C and delivered either sequentially or simultaneously. Conduction through the lesion area was measured. The lesions were analyzed histologically. At a perfusion rate of 0 mL/min, all RF lesions were transmural and without conduction. As little as 1 mL of flow through the artery during RF delivery could prevent transmural lesion formation by preserving a cuff of tissue along the length of the vessel. High-energy delivery (45 W) and very high tissue temperatures (93 degrees C) were needed to overcome the protective effect of vascular perfusion at 12 mL/min. The volume of preserved myocardium was related to arterial perfusion rate, artery diameter, and lesion temperature but not to the sequence of RF delivery (sequential versus simultaneous). Conduction persisted through the RF lesion in 20 experiments. Conduction through the lesion was related to the arterial perfusion rate and volume or cross-sectional area of preserved myocardium. Flow through even small intramyocardial vessels can prevent transmural lesion formation and preserve conduction through an RF lesion. These findings may represent an unrecognized obstacle to the creation of linear RF lesions in the clinical setting.Keywords
This publication has 7 references indexed in Scilit:
- Acute and Chronic Electrophysiologic Changes Surrounding Radiofrequency LesionsJournal of Cardiovascular Electrophysiology, 2002
- The Effect of Ablation Sequence and Duration on Lesion Shape Using Rapidly Pulsed Radiofrequency Energy Through Multiple ElectrodesJournal of Interventional Cardiac Electrophysiology, 2000
- The Effect of a Residual Isthmus of Surviving Tissue on Conduction after Linear Ablation in Atrial MyocardiumJournal of Interventional Cardiac Electrophysiology, 2000
- A new system for catheter ablation of atrial fibrillationThe American Journal of Cardiology, 1999
- Primary angioplasty and thrombolysis in acute myocardial infarction. One year angiographic follow-upJournal of the American College of Cardiology, 1998
- Right and Left Atrial Radiofrequency Catheter Therapy of Paroxysmal Atrial FibrillationJournal of Cardiovascular Electrophysiology, 1996
- Increasing catheter ablation lesion size by simultaneous application of radiofrequency current to two adjacent sitesAmerican Heart Journal, 1993