Ventricular Fibrillation Associated With Use of Electrocautery
- 21 October 1974
- journal article
- research article
- Published by American Medical Association (AMA) in JAMA
- Vol. 230 (3) , 432-435
- https://doi.org/10.1001/jama.1974.03240030050027
Abstract
WITH the introduction of many electrical instruments into the operating room, there are increased hazards.1The electrocautery has decreased operating time and blood loss, but in conjunction with monitoring equipment has caused burns. The purpose of this report is to describe a case of ventricular fibrillation that occurred during use of the electrosurgical unit in a patient being monitored by many instruments. We are unaware of a previous report of ventricular fibrillation associated with use of the electrosurgical unit. Report of a Case A 24-year-old woman was scheduled for elective surgical correction of an asymptomatic secundum-type atrial septal defect. Cardiac catheterization three days before operation was uneventful and demonstrated a calculated 4.2:1 left to right shunt with right ventricular enlargement and no evidence of pulmonary hypertension. On the morning of surgery, premedication was given intramuscularly: atropine sulfate, 0.6 mg; morphine sulfate, 10 mg; and diazepam, 10 mg, two hoursKeywords