Externally Induced Irreversible Runaway Pacemaker
- 1 September 1981
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 4 (5) , 578-580
- https://doi.org/10.1111/j.1540-8159.1981.tb06231.x
Abstract
An implanted ventricular demand pulse generator started firing at a rate of 1200 ppm during routine pacemaker follow-up utilizing the Vitatron VA 1000 pacemaker analyzer. This resulted in complete loss of capture; fortunately the patient had an intrinsic heart rate of 50 bpm and remained asymptomatic. Emergency replacement of the pulse generator resulted in re-establishment of effective ventricular pacing. An artificial origin of the impulses from the measuring equipment was excluded. To the best of our knowledge this is the first report of an externally induced irreversible runaway pacemaker.Keywords
This publication has 9 references indexed in Scilit:
- Preliminary Experience with the Helifix Electrode for Transvenous Atrial ImplantationPacing and Clinical Electrophysiology, 1979
- “Runaway” in a Modern Generation PacemakerPacing and Clinical Electrophysiology, 1979
- Runaway Temporary Pacemaker: Case Report of a “Runaway TemporaryPacing and Clinical Electrophysiology, 1978
- Ventricular Ectopic Rhythms due to Rapid Runaway PacemakerChest, 1978
- Runaway pacemaker in seven patientsThe Journal of Thoracic and Cardiovascular Surgery, 1975
- Successful Treatment of Pacemaker-Induced Ventricular FibrillationChest, 1974
- Runaway pacemaker rhythm after 36 hours of electrical inactivity.Heart, 1974
- Runaway Pacemaker with Demonstration of Supernormal Period of ExcitationChest, 1974
- Runaway Demand Pacemaker: Report, In Vitro Reproduction, and ReviewThe Annals of Thoracic Surgery, 1970