The effect of breath size and posture on calibration of the respiratory inductive plethysmograph by multiple linear regression
- 1 January 1989
- journal article
- research article
- Published by European Respiratory Society (ERS) in European Respiratory Journal
- Vol. 2 (1) , 71-77
- https://doi.org/10.1183/09031936.93.02010071
Abstract
The accuracy of the respiratory inductive plethysmograph (Respitrace) for estimation of lung volume changes during quiet breathing and vital capacity (VC) manoeuvres was evaluated using a variant of the multiple linear regression (MLR) technique. We applied this technique successively on quiet breathing, on the whole VC, and on each of the four quarters of the VC separately. This was carried out in six body positions. The best estimation of tidal volumes was obtained when calibration factors calculated during quiet breathing were used. The best estimation of VC was obtained when the calibration factors were adapted to the level of lung inflation. These results indicate that, using a single position MLR calibration method, the Respitrace measures tidal and VC mouth volumes very accurately. The accuracy of this MLR method for estimation of the rib cage and abdominal contributions was validated by comparison with isovolume calibration factors. Both techniques gave very similar results during tidal breathing. However, the MLR calibration factors may have no physiological meaning (i.e. for volume partitioning) when they are calculated from VC manoeuvres, in which more than two degrees of freedom are involved.This publication has 11 references indexed in Scilit:
- Estimates of ventilation from body surface measurements in unrestrained subjectsJournal of Applied Physiology, 1986
- Importance of Breath Size in Calibrating the Respiratory Inductive PlethysmographChest, 1986
- Scintigraphic evaluation of shape of lung and chest in upright and head-down postureJournal of Applied Physiology, 1986
- Accuracy of Respiratory Inductive Plethysmograph over Wide Range of Rib Cage and Abdominal Compartmental Contributions to Tidal Volume in Normal Subjects and in Patients with Chronic Obstructive Pulmonary DiseaseAmerican Review of Respiratory Disease, 1984
- Single-position calibration of the respiratory inductance plethysmographJournal of Applied Physiology, 1983
- POSTURAL CHANGES IN RIB CAGE AND ABDOMINAL VOLUME-MOTION COEFFICIENTS AND THEIR EFFECT ON THE CALIBRATION OF A RESPIRATORY INDUCTANCE PLETHYSMOGRAPHPublished by Elsevier ,1983
- EVALUATION OF THE RESPIRATORY INDUCTIVE PLETHYSMOGRAPH IN MAN1983
- VALIDATION OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY USING DIFFERENT CALIBRATION PROCEDURESPublished by Elsevier ,1982
- Computer-aided measurement of breath volume and time components using magnetometersJournal of Applied Physiology, 1978
- Measurement of the separate volume changes of rib cage and abdomen during breathingJournal of Applied Physiology, 1967